Abstract
Aims: To evaluate the performance of two matrix-assisted laser desorption/ionization-time-of-flight mass spectrometry platforms to identify molds isolated from clinical specimens. Methods: Fifty mold isolates were analyzed on Bruker Biotyper® and Vitek® MS platforms. Two Bruker Biotyper extraction protocols were assessed alongside the US FDA-approved extraction protocol for Vitek MS. Results: The Bruker Biotyper modifiedNIH-developed extraction protocol correctly identified more isolates than Bruker's protocol (56 vs 33%). For species in the manufacturers' databases, Vitek MS correctly identified 85% of isolates, with 8% misidentifications. The Bruker Biotyper identified 64%, with no misidentifications. For isolates not in the databases, the Bruker Biotyper did not misidentify any, and Vitek MS misidentified 36%. Conclusion: Both the Vitek MS and Bruker Biotyper accurately identified the fungal isolates, howeverVitek MS was more likely to misidentify isolates than theBruker Biotyper.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.