Abstract

Aims: To evaluate the performance of two matrix-assisted laser desorption/ionization-time-of-flight mass spectrometry platforms to identify molds isolated from clinical specimens. Methods: Fifty mold isolates were analyzed on Bruker Biotyper® and Vitek® MS platforms. Two Bruker Biotyper extraction protocols were assessed alongside the US FDA-approved extraction protocol for Vitek MS. Results: The Bruker Biotyper modifiedNIH-developed extraction protocol correctly identified more isolates than Bruker's protocol (56 vs 33%). For species in the manufacturers' databases, Vitek MS correctly identified 85% of isolates, with 8% misidentifications. The Bruker Biotyper identified 64%, with no misidentifications. For isolates not in the databases, the Bruker Biotyper did not misidentify any, and Vitek MS misidentified 36%. Conclusion: Both the Vitek MS and Bruker Biotyper accurately identified the fungal isolates, howeverVitek MS was more likely to misidentify isolates than theBruker Biotyper.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.