Abstract

ObjectiveTo determine if bicarbonate values from venous blood gas (VBG) and plasma chemistry samples provided agreement in determining the bicarbonate criteria for the diagnosis and/or resolution of diabetic ketoacidosis (DKA). MethodsA retrospective chart review of data from patients admitted to a tertiary care hospital with a diagnosis of DKA over a four year period was performed. Paired bicarbonate values from a VBG and chemistry panel, if drawn within 60 minutes of each other, were compared. ResultsAt the time of diagnosis of DKA, 197 paired bicarbonate values were available for analysis with the mean difference between the two methods of testing of 2.5 mmol/L. 16 of the 197 (8%) paired values were discordant in meeting criteria for diagnosis of DKA. At the time of resolution of DKA, 83 paired bicarbonate samples were compared. The mean difference was 2.3 mmol/L. 20 of the 83 (24%) paired bicarbonate values showed discordance with regards to meeting the bicarbonate criteria for resolution of DKA. ConclusionDiscordance between bicarbonate results from different analysis methods may lead to different determinations as to whether or not a patient meets the biochemical definition for diagnosis and resolution of DKA.