Abstract

A new radiographic technique was used to compare apical transportation in four Ni-Ti rotary instrumentation sequences. Mesiobuccal canals of 60 extracted mandibular molars were randomly divided into four groups. Groups 1 and 3 were instrumented by crown-down and groups 2 and 4 by step-back technique with 0.06 ProFiles series 29 to size 6. In groups 3 and 4 Greater Taper files were first used in a crown-down manner. The central axes of initial and final instruments were radiographically superimposed to measure loss of working length (WL) and transportation at 0, 0.5, 1, 3, and 5 mm from WL. ANOVA test showed no significant differences among groups regarding degree of transportation or loss of WL. Transportation was negatively correlated with radius of curvature at 0.5 and 5 mm from WL. The results indicate that the operational sequence of ProFiles or preinstrumentation with GT files has no effect on degree of transportation and loss of WL.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.