Abstract

Primary angular acceleration calibration system is developed by Changcheng Institute of Metrology and Measurement (CIMM) to generate angular vibration and shock, which are traceable to the International System of Units (SI). It can be used to calibrate angular transducers, i.e. angular accelerometer, angular velocity transducer, and rotational angle transducer. Two kinds of system are used in the measurement of angular movement, one is based on circular grating and scanning heads, another is based on laser interferometer with diffraction grating. This paper introduce the comparison results of the two measurement systems in the measurement of angular movement under sinusoidal and shock excitation. The results of the investigations show a good accordance of the newly developed method of using grating and scanning heads measuring angular acceleration in comparison with the laser interferometer method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.