Abstract

Pulse shape discrimination (PSD) performance of two optimized PSD systems (one digital and one analog) is compared in this work. One system uses digital charge integration, while the other system uses analog zero crossing. Measurements were conducted with each PSD system using the CAEN V1720 (250MHz) data acquisition system. An organic-liquid scintillator, coupled to a photo-multiplier tube, was used to detect neutrons and gamma rays from a Cf-252 spontaneous-fission source. The PSD performance of both systems was optimized and quantified using a traditional figure-of-merit (FOM) approach. FOM's were found for three, 300 keVee light-output bins, spanning from 100 to 1000keVee, and one larger bin from 100 to 1800keVee. Digital PSD outperformed analog PSD in the lowest light-output bin by approximately 50%, and by 11% for the highest light-output bin.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.