The aim of this study was to compare the amount of residual organic debris on endodontic instruments that had been cleaned in either an ultrasonic bath or a washer disinfector prior to sterilisation. A total of 90 endodontic files of varying sizes were used to clean and shape root canals in extracted teeth and were then placed in endodontic file holders. Test group 1 (36 files) were ultrasonically cleaned for 10 min and test group 2 (36 files) were cleaned in a washer disinfector. A control group (18 files) were not cleaned at all. Following sterilisation, all the files were visually inspected under a light microscope and scored using an established scale. The results showed that both test groups had significantly less residual debris than the control group. Comparing the test groups, the files that had been cleaned ultrasonically had significantly less debris than those cleaned in the washer disinfector. The design of the instrument holder may have been a factor in this result. More research is needed into the use of washer disinfectors in the cleaning of small dental instruments that have a complex surface structure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.