Abstract

The emergence of a novel coronavirus, namely, SARS-CoV-2, necessitated the use of rapid, accurate diagnostics to quickly diagnose COVID-19. This need has increased with the emergence of new variants and continued waves of COVID-19 cases. The ID NOW COVID-19 assay is a rapid nucleic acid amplification test (NAAT) that is used by hospitals, urgent care facilities, medical clinics, and public health laboratories for rapid molecular SARS-CoV-2 testing at the point of care. The District of Columbia Department of Forensic Sciences Public Health Laboratory Division (DC DFS PHL) implemented ID NOW COVID-19 testing in nontraditional laboratory settings, including a mobile testing unit, health clinic, and emergency department, to assist with rapid identification and isolation for populations at high risk of SARS-CoV-2 transmission in the District of Columbia. The DC DFS PHL provided these nontraditional laboratories with safety risk assessment, assay training, competency assessment, and quality control monitoring as parts of a comprehensive quality management system (QMS). We assessed the accuracy of the ID NOW COVID-19 assay when operated in the context of these trainings and systems. This was done by comparing results from 9,518 paired tests, and strong agreement (κ = 0.88, OPA = 98.3%) was found between the ID NOW COVID-19 assay and laboratory-based NAATs. These findings indicate that the ID NOW COVID-19 assay can be used to detect SARS-CoV-2 in nontraditional laboratory settings when used within the context of a comprehensive QMS.