Abstract

An anthropomorphic airway tree phantom was imaged with both hyperpolarized (HP) 3He MRI using a dynamic projection scan and computed tomography (CT). Airway diameter measurements from the HP 3He MR images obtained using a newly developed model-based algorithm were compared against their corresponding CT values quantified with a well-established method. Of the 45 airway segments that could be evaluated with CT, only 14 airway segments (31%) could be evaluated using HP 3He MRI. No airway segments smaller than approximately 4 mm in diameter and distal to the fourth generation were adequate for analysis in MRI. For the 14 airway segments measured, only two airway segments yielded a non-equivalent comparison between the two imaging modalities, while eight more had inconclusive comparison results, leaving only four airway segments (29%) that satisfied the designed equivalence criteria. Some of the potential problems in airway diameter quantification described in the formulation of the model-based algorithm were observed in this study. These results suggest that dynamic projection HP 3He MRI may have limited utility for measuring airway segment diameters, particularly those of the central airways.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.