Abstract

An indirect comparison has been made in the air kerma standards for high dose rate (HDR) 192Ir brachytherapy sources at the Laboratoire National Henri Becquerel (LNHB) and the National Physical Laboratory (NPL). The measurements were carried out at both laboratories between November and December 2004. The comparison was based on measurements using well-type transfer ionization chambers and two different source types, Nucletron microSelectron HDR Classic and version 2. The results show the reported calibration coefficients to agree within 0.47% to 0.63%, which is within the overall standard uncertainty of 0.65% reported by both laboratories at the time of this comparison. Following this comparison, some of the NPL primary standard correction factors were re-evaluated resulting in a change of +0.17% in the overall correction factor. The new factor was implemented in May 2006. Applying the revised chamber factor to the measurements reported in this comparison report will reduce the difference between the two standards by 0.17%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.