Abstract

This study compared the performance of 3 automated immunoassays, Architect® (Abbott), Immulite® (Siemens) and Liaison® (Diasorin), for Epstein–Barr virus (EBV) serology. Ninety-one serum samples collected in Amiens University Hospital were analyzed for the presence of Viral Capsid Antigen (VCA) IgG and IgM and Epstein–Barr Nuclear Antigen (EBNA) IgG. The agreement between the 3 assays was calculated for each marker individually and for determination of the EBV profile, based on interpretation of the combination of these 3 EBV markers.Although similar results were obtained with Architect® and Liaison®, several discordant results were observed with Immulite®, particularly for EBNA IgG. A large number of EBNA IgG-positive results were observed, which interfered with interpretation of the EBV profile. In contrast, Immulite® performed similarly to the 2 other assays for detection of VCA IgM.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.