Abstract

A unique preload responsive pulsatile pump was compared to a centrifugal pump in total cardiac support in 25-kg canines (n = 6, each group) in the left atrial-to-aorta mode during 5 h of ventricular fibrillation. With steady flow, there was immediate drop in output from 2.1 +/- 1.0 L/min to 1.4 +/- 0.3 L/min, followed by further reduction to 0.9 +/- 0.2 L/min during 5 h of ventricular fibrillation. With a pulsatile pump, there was no significant reduction from control of 2.4 +/- 0.6 L/min and no decline during 5 h of ventricular fibrillation. With steady flow, systemic vascular resistance (SVR) rose significantly from 1,762 dyne-s-cm-5 immediately on pump to 3,013 dyne-s-cm-5 at 5 h. With physiologic pulsatile flow, significant elevation of SVR did not occur. When stressed, due to diminished left atrial return, the centrifugal pump displayed line chatter and streaks of microbubbles, whereas the pulsatile pump did not. Crystalloid volume replacement with the centrifugal pump was 6.5 +/- 1.9 L, and with the preload responsive pulsatile pump, 5.6 +/- 1.3 L. It is concluded that in the left atrial-to-aorta mode during 5 h of ventricular fibrillation and with comparable volume replacement, total cardiac support of canines is associated with lower SVR with physiologic pulsatile flow and is not accompanied by line chatter and cavitation with this preload responsive pump.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.