Abstract

ObjectivesProcalcitonin (PCT) is an important biomarker of sepsis and respiratory infections. Various automated immunoassays for measuring PCT in patient plasma are available in medical laboratories. However, due to a lack of international reference material for PCT, the assays are not always comparable. Design and methodsIn this study, we compared a new turbidimetric immunoassay from DiaSys, measured on the Abbott Architect c16000 and Alinity c, with four BRAHMS-associated chemiluminescence immunoassays (Abbott Architect i2000SR, Alinity i, Roche Cobas e411 and DiaSorin Liaison XL) using 120 random patient plasma samples from the clinical laboratory routine at the University Medical Center Goettingen. ResultsThe DiaSys assay showed clear differences as compared to the BRAHMS-associated assays when measured on Architect c: i.e. 58% positive mean bias vs. Architect i, 67% vs. Cobas and 23% vs. Liaison. As a result, additional 19% our patients would have a suspected bacterial infection, when using PCT values from the DiaSys assay and commonly accepted decision limits. A crosscheck of the DiaSys calibrator on the BRAHMS-associated systems showed a low recovery of the calibrator material (approx. 50%). ConclusionsOverall, this study shows significant differences between the DiaSys and BRAHMS-associated assays. This could be attributed to a potential DiaSys calibrator problem. This highlights the need for an international reference material for harmonization of the PCT assays.