Abstract

To compare the performance of a handheld fundus camera with standard 50° visual field to ultra-widefield (UWF) table-top fundus camera in diabetic retinopathy (DR) detection and grading. Patients affected by diabetes mellitus and referred to our diabetic retinopathy clinic were enroled and underwent fundus photography in mydriasis. All photos were taken using the ultra-widefield table-top fundus camera Zeiss Clarus™ 500 (four fields per eye) and the Optomed Aurora® handheld fundus camera (3 fields per eye). The following parameters were analysed: the gradability of the images, the grade of DR, and diabetic maculopathy (DM), the presence of hypertensive retinopathy (HR) and the presence of other ocular diseases. We enroled 759 eyes of 384 diabetic patients and analysed 5313 fundus photos. The handheld fundus camera obtained a sensitivity of 84.2% and specificity of 95.4% for referable cases. Moreover, it obtained, compared to UWF, an almost perfect agreement with linear weighting for DR, DM and HR (k = 0.877, k = 0.854, and k = 0.961, respectively). The lowest sensitivity was achieved for proliferative DR (58.7% sensitivity, 100% specificity). Optomed Aurora® handheld fundus camera imaging showed a strong agreement compared to UWF in grading DR, considering all DR and DM grades, in mydriasis. However, the use of UWF imaging increases the detection of referable eyes.