BackgroundCardiac arrhythmias are remarkably common and routinely go undiagnosed because they are often transient and asymptomatic. Effective diagnosis and treatment can substantially reduce the morbidity and mortality associated with cardiac arrhythmias. The Zio Patch (iRhythm Technologies, Inc, San Francisco, Calif) is a novel, single-lead electrocardiographic (ECG), lightweight, Food and Drug Administration–cleared, continuously recording ambulatory adhesive patch monitor suitable for detecting cardiac arrhythmias in patients referred for ambulatory ECG monitoring. MethodsA total of 146 patients referred for evaluation of cardiac arrhythmia underwent simultaneous ambulatory ECG recording with a conventional 24-hour Holter monitor and a 14-day adhesive patch monitor. The primary outcome of the study was to compare the detection arrhythmia events over total wear time for both devices. Arrhythmia events were defined as detection of any 1 of 6 arrhythmias, including supraventricular tachycardia, atrial fibrillation/flutter, pause greater than 3 seconds, atrioventricular block, ventricular tachycardia, or polymorphic ventricular tachycardia/ventricular fibrillation. McNemar's tests were used to compare the matched pairs of data from the Holter and the adhesive patch monitor. ResultsOver the total wear time of both devices, the adhesive patch monitor detected 96 arrhythmia events compared with 61 arrhythmia events by the Holter monitor (P < .001). ConclusionsOver the total wear time of both devices, the adhesive patch monitor detected more events than the Holter monitor. Prolonged duration monitoring for detection of arrhythmia events using single-lead, less-obtrusive, adhesive-patch monitoring platforms could replace conventional Holter monitoring in patients referred for ambulatory ECG monitoring.

Highlights

  • Cardiac arrhythmias are remarkably common and routinely go undiagnosed because they are often transient and asymptomatic

  • We aimed to evaluate the diagnostic utility of a novel adhesive patch monitor for up to 14 days compared with standard 24-hour Holter monitoring

  • Between April 2012 and July 2012, patients referred to the cardiac investigations laboratory at Scripps Green Hospital (La Jolla, Calif) for ambulatory ECG monitoring were fitted with an adhesive patch monitor and a 24-hour Holter monitor

Read more

Summary

Introduction

Cardiac arrhythmias are remarkably common and routinely go undiagnosed because they are often transient and asymptomatic. METHODS—A total of 146 patients referred for evaluation of cardiac arrhythmia underwent simultaneous ambulatory ECG recording with a conventional 24-hour Holter monitor and a 14day adhesive patch monitor. The primary outcome of the study was to compare the detection arrhythmia events over total wear time for both devices. RESULTS—Over the total wear time of both devices, the adhesive patch monitor detected 96 arrhythmia events compared with 61 arrhythmia events by the Holter monitor (P < .001). CONCLUSIONS—Over the total wear time of both devices, the adhesive patch monitor detected more events than the Holter monitor. Prolonged duration monitoring for detection of arrhythmia events using single-lead, less-obtrusive, adhesive-patch monitoring platforms could replace conventional Holter monitoring in patients referred for ambulatory ECG monitoring

Objectives
Methods
Results
Conclusion