Abstract
Aims: To compare the performance of the four latest models of glucose meters in capillary blood glucose monitoring during pregnancy. Methods: 208 pregnant women with gestational diabetes were recruited. Each subject had simultaneous capillary glucose monitoring by two study glucose meters and venous plasma glucose assay. The performance of four glucose meters was compared using error grid analysis (EGA) and the agreement between the meter readings and plasma glucose by Bland-Altman plot analysis. Results: Elite, Advantage II and CareSens had more than 90% of readings in the acceptable target range of EGA. CareSens had the lowest mean bias by Bland-Altman analysis while Advantage II had the highest proportion of readings within 5% difference from plasma glucose. Readings from all glucose meters except Optium were not influenced by the change in maternal hematocrit levels. Conclusions: The performance of four study glucose meters appeared very similar.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.