Abstract
The full energy peak efficiency (ɛ) relation of HpGe detector is determined and compared with the "spectrum shape" relation of natural background (BG) for the (121.8-1408) keV energy range. A one liter Marinelli beaker soil standard containing the Eu-152 material is used in calibration and measurement. A MATLAB fitting program were used obtain the efficiency relation from the experimental values, and to obtain the BG "spectrum shape" relation from the count rates at the corresponding energy values used. The ɛ / BG factor values were 1.154 at the 121.8keV and increase with energy till reaching a maximum value of 4.358 at about 778keV energy and then decrease. These values may be used to extract efficiency from the BG values at the specified measurement conditions. The obtained results may participate in extending our understanding of the standardization concept in radiation measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.