Abstract

Introduction: The aim of this study is evaluation of two Quality Assurance methods in sliding window IMRT technique and Determination of gamma index in both methods Materials and Methods: In this study two tools named phantom and software has been used to perform QA on the treatment plan before the actual treatment by Varian linear accelerator Clinac 600C. For this purpose, 20 patient’s treatment plan were assessed with the 3% and 3mm criteria. Results: Each treatment field demonstrated that the gamma index in the delta4 phantom was more than the Epiqa software for all patients. The most uncertainty for a treatment plan in delta4 phantom was 0.88% and for Epiqa software was 1.53%. (p- value <0.05) Conclusion: In spite of these variations in results and also studying similar researches, these tools can be used as a reliable QA method in radiotherapy centers. Delta4 phantom is preferred to Epiqa software because it is a dosimeter. However, Epiqa software is economical to install. therefore can be used for those centers which cannot afford delta4 phantom due to economic consideration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.