Abstract

A 177Lu primary standard was developed at the ENEA-INMRI in the frame of an international comparison organized by BIPM and piloted by NIST (USA). The CIEMAT/NIST method with 3H standard source as tracer was used for standardizing a solution of 177Lu. The activity value was compared also with the measurements of the same mother solution carried out by the 4πγ integral counting method. Particular efforts were made to identify and quantify the long–lived 177mLu impurity in the mother solution. The results obtained by the two methods are in good agreement within their standard uncertainties. The arithmetic mean of the two values is in good agreement with the Comparison Reference Value (CRV). By the new primary standard two well-type ionization chambers (ICs), one fixed and the other one portable, were calibrated with an uncertainty lower than 2%. These ICs are used for routinely applications in the activity measurements of short-lived radionuclides particularly useful for medical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.