Abstract

In high-intensity focused ultrasound treatment, targeting its focal region prior to the thermal coagulation is required for the safety and efficacy of the treatment because the size and position of the thermally effective focal zone depend on the acoustic properties of the tissue. Acoustic radiation force impulse imaging can estimate the focal region from the displacement caused by ARF which stems from ultrasonic attenuation. Thermal strain (TS) imaging is another targeting method, which is based on temperature rise caused by tissue absorption. In this study, the TS imaging and temperature measurement by thermocouple were conducted for the same location, and a linear relationship between the TS and temperature rise was confirmed. In addition, the differences of the distribution depending upon the physical phenomena such as the thermo-acoustic lens effect and the effect of acoustic reflection were observed by comparing two imaging methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.