Abstract

State-of-the art dimensional metrology was used to measure the throat diameter and throat curvature of nine critical flow venturis (CFVs) with nominal throat diameters ranging from 5mmto25mm. The throat curvature was used in calculating the theoretical discharge coefficients, while the throat diameter was used in computing the experimental discharge coefficients. The nine CFVs were calibrated in dry air using two NIST primary flow standards with expanded uncertainties of 0.05% and 0.09%, respectively. The calibration data span a Reynolds number range from 7.2×104 to 2.5×106, including laminar, transition, and turbulent flow regimes. By correcting for both the throat diameter and curvature, the agreement between predicted and measured discharge coefficients was less than 0.17% in the turbulent regime and less than 0.07% in the laminar regime.