Abstract

Gold standard for pulse-wave velocity (PWV) measurement is determination of the carotid-femoral cfPWV, reflecting regional aortic PWV. Nevertheless, in several echocardiographic laboratories, PWV is measured locally, most commonly at the common carotid artery (CCA). The aim of this study was to compare regional and local PWV values in healthy volunteers. The study population consisted of 22 prospectively enrolled healthy subjects, mean age 38.7 ± 11.1 years, 50% male. For regional PWV measurement, we evaluated cfPWV with a standard echo scanner. Regional PWV was measured at the CCA, with semiautomated dedicated software (MyLab, EsaOte, Italy). cfPWV and local PWV values correlated significantly with high Pearson correlation coefficient (0.62, P = 0.002). Mean regional cfPWV (9.29 ± 3.73 m/s), however, was significantly higher than mean local PWV value (5.96 ± 1.08 m/s) (P < 0.001). The difference persisted in the subgroup analysis using different cfPWV cutoff values (10, 9, 8, and 7 m/s), except for subjects with cfPWV ≤7 m/s, where regional and local PWV values were similar. In a group of healthy volunteers, regional and local PWV values showed a good correlation. However, regional PWV was significantly higher than local PWV. These findings should be carefully taken into account when using this technique in the clinical setting.