Abstract

The properties of the maximum likelihood with expectation maximization (ML-EM) and the weighted least squares with conjugate gradient (WLS-CG) algorithms for use in compensation for attenuation and detector response in cardiac SPECT imaging were studied. A realistic phantom, derived from a patient X-ray CT study to simulate /sup 201/Tl SPECT data, was used in the investigation. In general, the convergence rate of the WLS-CG algorithm is about ten times that of the ML-EM algorithm. Also, the WLS-CG exhibits a faster increase in image noise at large iteration numbers than the ML-EM algorithm. >

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.