Abstract
Differential leukocyte (WBC) counts in blood from clinically healthy silver foxes (n=32) and blue foxes (n=37) obtained from an automated hematology analyzer (Technicon H*1 Hematology System) with canine software were compared with microscopic differential WBC counts (M-diff). There was good agreement between the automated differential cell count (A-diff) and the M-diff for neutrophil and lymphocyte percentages. The correlation was lower for monocyte percentages and variable for eosinophil percentages. There was no significant difference between the A-diff and M-diff in either fox species. The A-diff counts were very precise, and may be a good alternative to the traditional M-diff for screening populations of clinically healthy foxes or for studies on stress and animal welfare. Intercept values, however, indicated a constant bias that must be taken into account before interpreting results based on different methods of analysis
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.