Abstract
The Intermediary Checking of the calibration the electrometers, to ensure that the calibration factor of a given instrument is valid for a longer time, has been developed at LNMRI-IRD since 2019. It works with three of the methods for generating electric current of small values (pA): capacitance, resistance (Ohm) and ionization current (radiation sources). In this article, we will evaluate the performance of electrometers using the ionization current method and we will make a comparison between the results obtained by this method with the results that uses the Ohm method. From this comparison, we could see that the method using radiation sources had a better performance of the electrometers, however, in terms of cost-benefit, the Ohm method better serves the intermediary checking of the calibration the electrometers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.