Abstract

Glow discharge mass spectrometry (GDMS) using either a direct current (dc) or a radiofrequency (rf) discharge has been evaluated. For conducting samples, comparable results are obtained, as indicated by the uniformity in relative sensitivity factors (RSFs). For the analysis of rare earth oxide materials, the RSFs for dc GDMS are far less affected by changes in discharge pressure and sample to exit-orifice distance than those obtained for rf GDMS. Despite the higher variability of RSFs for rf GDMS, very stable and reproducible plasma conditions are obtained when working at a fixed pressure and distance. The latter is indicated by the internal (within-sample) and external (sample-to-sample) reproducibility of less than 5 and 15%, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.