Abstract

Objective. To determine the diagnostic accuracy of placental alpha microglobulin-1 assay and standard diagnostic methods for detecting rupture of membrane. Study Design. Prospective diagnostic study, between June 2011 to November 2011 at a tertiary centre. Initial evaluation included both the standard diagnostic methods for rupture of membranes and placental alpha microglobulin-1 immunoassay. The actual rupture of membranes was diagnosed on review of the medical records after delivery (absence of membrane or a positive pad chart). Main Outcome Measures. Placental alpha microglobulin-1 immunoassay and standard diagnostic methods for diagnosis of rupture of membrane. Results. A total of 211 patients were recruited. At initial presentation, 187 patients (88.6%) had ruptured membranes, while 24 patients (11.4%) had intact membranes. Placental alpha microglobulin-1 immunoassay confirmed rupture of membranes at initial presentation with a sensitivity of 95.7% (179 of 187), specificity of 100% (24 of 24), positive predictive value of 100% (179 of 179), and negative predictive value of 75.0% (24 of 32). By comparison, the conventional standard diagnostic methods had a sensitivity of 78.1% (146 of 187), specificity of 100% (24 of 24), positive predictive value of 100% (146 of 146), and negative predictive value of 36.9% (24 of 65) in diagnosing rupture of membrane. Conclusion. Placental alpha-microglobulin-1 immunoassay is a rapid and accurate method for confirming the diagnosis of rupture of membrane. It was superior to conventional standard diagnostic methods (pooling, nitrazine, and ferning), the nitrazine test alone or fern test alone.

Highlights

  • Prelabour rupture of the membranes (PROM) is defined as rupture of the fetal membranes before the onset of labour

  • The PMAG-1 immunoassay alone has an overall sensitivity of 97.5%, specificity of 100%, positive predictive value (PPV) of 100%, and negative predictive value (NPV) of 75.0%

  • The AmniSure immunoassay involves detecting the presence of high concentration of glycoprotein PAMG-1 in the cervicovaginal secretion

Read more

Summary

Objective

To determine the diagnostic accuracy of placental alpha microglobulin-1 assay and standard diagnostic methods for detecting rupture of membrane. Prospective diagnostic study, between June 2011 to November 2011 at a tertiary centre Initial evaluation included both the standard diagnostic methods for rupture of membranes and placental alpha microglobulin immunoassay. Placental alpha microglobulin-1 immunoassay and standard diagnostic methods for diagnosis of rupture of membrane. Placental alpha microglobulin-1 immunoassay confirmed rupture of membranes at initial presentation with a sensitivity of 95.7% (179 of 187), specificity of 100% (24 of 24), positive predictive value of 100% (179 of 179), and negative predictive value of 75.0% (24 of 32). The conventional standard diagnostic methods had a sensitivity of 78.1% (146 of 187), specificity of 100% (24 of 24), positive predictive value of 100% (146 of 146), and negative predictive value of 36.9% (24 of 65) in diagnosing rupture of membrane. It was superior to conventional standard diagnostic methods (pooling, nitrazine, and ferning), the nitrazine test alone or fern test alone

Introduction
Materials and Methods
Result
Discussion
Limitation
Conclusion
Findings
Conflict of Interests
Disclosure