Abstract
To investigate the high-accuracy roundness metrology of a needle roller 1.5 mm in diameter and 5.8 mm in length using the stitching linear scan method, a ruby ball stylus with a tip radius of 150 μm and a diamond stylus with a tip radius of 2 μm were employed to perform experiments under the same conditions. The precision coordinate data, derived from the needle roller's cross-sectional circumference, were segmented into uniform eighths, each scanned with the stylus of a roughness measuring machine. The roundness profile of the needle roller was obtained by stitching the arc profiles, which were characterized according to the precision coordinate data of the arcs. The cross-correlation function, Euclidean distance, residual sum of squares, position error, and curvature of the measured arcs were used to evaluate the results, which can reflect the performance of the stylus. A comparison of the results obtained using the ruby ball stylus versus the diamond stylus demonstrates the ruby ball stylus' greater suitability for use in the roundness metrology of the needle roller bearing examined in this paper.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.