Abstract
This study aimed to evaluate the influence of various sterilization techniques on the cyclic fatigue resistance, bending strength, and torsional fatigue resistance of Tornado Blue and Race Evo rotary endodontic file systems. Both the file systems underwent a series of 15 cycles of sterilization according to the following study groups: control group (G1; i.e., without treatment), chemical sterilization with 5.25% sodium hypochlorite (G2), dry heat sterilization (G3), glass beads sterilization (G4), and steam autoclaving (G5). Each group consisted of 8 samples (n = 8/group). Statistical analysis was performed using a two-way ANOVA and Tukey’s tests (p ≤ 0.05). Mean values for the number of cycles between both files were 1291 ± 52.79 rpm and 2095 ± 52.91 rpm for Tornado Blue and Race Evo, respectively. The torsional fatigue resistance was 8.00 ± 0.85 NCF and 10.70 ± 0.85 NCF and the bending strength was 486.94 ± 20.97 gf and 564.17 ± 20.97 gf for Tornado Blue and Race Evo, respectively. The results showed no significant difference in the mechanical properties followed by different sterilization protocols. However, a significant difference was observed in tested properties between the two file systems (p ≤ 0.05). The outcome is reassuring for clinicians that the essential process of routine sterilization at clinics seems safe with regards to mechanical integrity of the endodontic files.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.