Abstract

BackgroundThe pressure cooker (PC) technique uses a second microcatheter to produce a proximal occlusion to prevent the reflux of liquid embolic agent (LEA) injected through a more distal microcatheter, and can be used to treat arteriovenous malformations and dural arteriovenous fistulae. The liquid embolic Magic Glue (MG) (N-hexyl cyanoacrylate (NHCA)) has been introduced as an alternative to N-butyl cyanoacrylate (NBCA). Our goals were to compare the extent of embolization of rete mirabile with or without the PC technique using NBCA or MG, and to compare the proximal occlusions obtained with MG or NBCA while using the PC technique in a renal arterial model. MethodsRete mirabile were embolized with (n = 4) and without (n = 4) the PC technique, using MG (n = 4) or NBCA (n = 4). A renal arterial model was then used to study the characteristics of the MG plug (n = 10) used for the PC technique, and resistance to catheter withdrawal as compared to NBCA (n = 4). Specimens were analyzed macro- and microscopically and compared to angiographic results. ResultsExtent of rete embolization with CYA agents was not significantly greater when using the PC technique. Results were similar with both types of cyanoacrylate (p = 0.657). The force necessary to withdraw the microcatheter was less with MG than with NBCA (p = 0.035). ConclusionMG was similar to NBCA in extent of rete embolization. Less traction force was necessary to withdraw trapped non-detachable microcatheters using MG compared to NBCA.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call