Abstract

Muscle-recorded transcranial electrical stimulation motor-evoked potentials (mTc-MEPs) are used to assess the spinal cord integrity. They are commonly recorded with subcutaneous needle or surface electrodes, but the different characteristics of mTc-MEP signals recorded with the two types of electrodes have not been formally compared yet. In this study, mTc-MEPs were simultaneously recorded from the tibialis anterior (TA) muscles using surface and subcutaneous needle electrodes in 242 consecutive patients. Elicitability, motor thresholds, amplitude, area under the curve (AUC), signal-to-noise ratio (SNR), and the variability between mTc-MEP amplitudes were compared. Whereas amplitude and AUC were significantly higher in subcutaneous needle recordings (p < 0.01), motor thresholds and elicitability were similar for surface and subcutaneous needle recordings. Moreover, the SNRs were >2 in more than 99.5% of the surface and subcutaneous needle recordings, and the variability between consecutive amplitudes was not significantly different between the two recording electrode types (p = 0.34). Surface electrodes appear to be a good alternative to needle electrodes for spinal cord monitoring. They are non-invasive, can record signals at similar threshold intensities, have adequately high SNRs, and record signals with equivalent variability. Whether surface electrodes are non-inferior to subcutaneous needle electrodes in detecting motor warnings is investigated in part II of the NERFACE study.