Abstract

Spectrally resolved UV measurements are important for the study of biologically relevant UV in relation to changes in atmospheric parameters. The inter‐comparison of spectral instruments is essential as measurement techniques and calibrations are not standardized. The differences in slit functions cause large spectral variations when comparing the spectral readings directly. The method described, which compares spectral readings using different instruments, corrects for differences of wavelength calibrations and slit functions, and does not require knowledge of additional atmospheric parameters and UV‐transfer model calculations. The wavelength alignment has an accuracy of 0.02 nm over the wavelength interval from 300–400 nm, and a reproducibility of 0.01 nm. The robustness of the methods and reproducibility of results are shown in the evaluation of a seven day intercomparison campaign with three different scanning spectroradiometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call