The three different ligands (Q 2 to Q 4) based on 2-amino-1,4-naphthoquinone (Q 1), have been synthesized and explored as neutral ionophores for preparing polyvinyl chloride-based membrane sensors selective to indium (III). The addition of potassium tetrakis(4-chlorophenyl) borate and various plasticizers, viz., o-NPOE, DBP, DBBP, DOP and CN has been found to substantially improve the performance of the sensors. The best performance was obtained with the sensor no. 16 having membrane of ligand (Q 2) with composition (%, w/w) ionophore Q 2 (3.0%):PVC (30.0%): o-NPOE (63.0%):KTpClPB (4.0%). This sensor exhibits Nernstian response with slope 19.8 mV/decade of activity in the concentration range 2.5 × 10 −7 to 1.0 × 10 −2 M indium (III), performs satisfactorily over wide pH range of (2.5–7.5) with a fast response time (10 s). The sensor was also found to work satisfactorily in partially non-aqueous media up to 20% (v/v) content of acetonitrile, ethanol and methanol. The proposed sensor can be used over a period of 3.5 months without significant drift in potentials. The quantitative application of sensor was also evaluated by comparative analysis of artificially made sea water with AAS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.