Abstract
N +p and p +n diodes were fabricated on p- and n-type 6H–SiC substrates with epitaxial layers, respectively. The charge induced in the diodes by 9 MeV oxygen (O) and nickel (Ni) ions was measured using Transient Ion Beam Induced Current (TIBIC) to clarify the capability of these diodes as particle detectors. As a result of the TIBIC measurements using 9 MeV O, the Charge Collection Efficiency (CCE) of around 83% was obtained for both p +n and n +p diodes. Since the CCE value includes the consumption of incident ion energy in an Al electrode and the n + (p +) region as well as the decay of charge in the measurement system, the CCE value obtained in this study indicates that SiC n +p as well as p +n diodes are suitable for particle detectors. On the other hand, in the case of 9 MeV Ni ion irradiation, the CCE for both n +p and p +n diodes decreases due to the Auger recombination in dense electron–hole pairs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.