The surface roughness of polished glass substrates as well as metal an dielectric coatings is studied using atomic force microscopy, hard and soft x-ray scattering at the wavelength (lambda) equals 0.154 nm and 4.47 nm, and light scattering (LS) at (lambda) equals 325 nm. It is demonstrated that all the methods, permitting the determination of PSD functions in partly overlapping, partly different ranges of spatial frequency, are in a good agreement in spite of different physical principles underlying the methods. The possible reasons for some differences in the PSD functions determined form different measurements are discussed. The main of them are a more difficult interpretation of LS data when dielectric coatings are present and a limitation on the angular range of hard x-ray scattering measurements imposed by the applicability of the perturbation theory used for experimental data processing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.