Abstract

The photovoltaic method for the detection of liquid-phase photoionization in the absence of an applied electric field is compared to the traditional photoconductive method. Each system is characterized by means of signal and signal-to-noise ratio measurements with respect to laser pulse energy, applied voltage, electrode separation distance, solute concentration, and solvent composition. Although both techniques exhibit similar detection limits (1 × 10−7 M) for solutes in polar solvents such as methanol and water, the photovoltaic technique has a more extensive linear dynamic range (103). Moreover, this technique allows detection of solutes in ionic solutions with comparable detectability and linearity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.