Abstract

AIM: The study aims to compare the results of objective parameters such as autorefractometers of the open field Grand Seiko and closed field TONOREF III. and the subjective parameters such as the positive of relative accommodation (PRA) and the amplitude of accommodation (AA).
 MATERIAL AND METHODS: 30 children (60 eyes) with low and moderate myopia (on average -2.96 0.17 D) aged from 8 to 12 years (on average 10.04 0.24 years) were examined. Subjective (PRA, AA) and objective parameters of binocular adaptation (BAO) and monocular adaptation (MAO) response on the Grand Seiko Binocular Open Field Autorefkeratometer WR - 5500K (Japan) and the AA on the automatic refractokeratotonometer pakhimetre TONOREF III (Nidek, Japan) were determined.
 RESULTS: The average of BAO and MAO at 33 cm was -1.93 0.04 D and 1.86 0.05 D, respectively. The average PRA was 1.5 0.16 D. The objectively measured average AA was 5.25 0.4 D. The average minimum AA value was -2.86 0.16 D, and the average maximum value was 8.11 0.46 D. The subjective AA on the Iksar device was on average 4.17 0.43 D; Amin, on average -3.77 0.26 D; Amax, on average was -7.94 0.59 D.
 CONCLUSION: The objective and subjective measurements of AA produced comparable results. BAO and MAO reflected other characteristics of accommodation, different from its amplitude, and characterized the adequacy of the accommodation response to a specific accommodation task. The advantage of objective accommodation is that it is independent of the patients responses and intellectual level.