Abstract

To compare image quality of turbo spin-echo (TSE) with BLADE [which is also named periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER)] on magnetic resonance imaging (MRI) for upper abdomen. This study involved the retrospective evaluation of 103 patients (63 males, 40 females; age range 19-76years; median age 53.8years) who underwent 3.0T MRI with both conventional TSE T2-weighted imaging (T2WI) and BLADE TSE T2WI. Two radiologists assessed respiratory motion, gastrointestinal peristalsis, and vascular pulsation artifacts, as well as the sharpness of the liver and pancreas edges. Scores for all magnetic resonance (MR) images were recorded. Wilcoxon's rank test was used to compare hierarchical data. Cohen's kappa coefficient was adopted to analyze interobserver consistency. Compared to TSE T2WI, BLADE TSE T2WI reduced all of the examined motion artifacts and increased the sharpness of the liver and pancreas edges (all P<0.05). Medium to good interobserver consistency was obtained for evaluating these indicators. The scanning time of BLADE TSE T2WI was 4-16s shorter than that of conventional TSE T2WI. Compared to TSE sequence, the BLADE technique can reduce the respiratory motion, gastrointestinal peristalsis, and vascular pulsation artifacts, while decreasing the scanning time and improving the anatomic detail and image quality.