Abstract

The accurate assessment of cardiac function in mice is challenging because of their small heart size and rapid heart rate. We examined the usefulness of novel high-resolution echocardiography (HRE) with a 30-MHz transducer in evaluating cardiac function in 20 mice compared with conventional echocardiography (CE) with a 13-MHz transducer. The left ventricular (LV) regional wall motion (RWM), LV end-diastolic dimension, fractional shortening, anterior LV wall thickness, E/A, and myocardial performance index were assessed. RWM analysis was more feasible by HRE than by CE (P < .05). Interobserver agreement in RWM analysis and correlation in LV end-diastolic dimension, fractional shortening, anterior LV wall thickness, E/A, and myocardial performance index were all better with HRE than CE. HRE is superior to CE in assessing LV function in mice. HRE is potentially a useful method for accurate assessment of cardiac function in various mice models.