Abstract

During 1990-1992 a study of dose and image quality in mammography was carried out in the Madrid area. Values of air kerma and mean glandular dose per film estimated at 14 public and private centres (17 X-ray units) by means of the Leeds TOR(MAX) mammographic phantom and from a sample of patients were reported in a previous paper. This second paper describes image quality results obtained by the same Leeds TOR(MAX) phantom. Inferior image quality was found for units without a grid (most of them older systems) and/or using aluminium filtration. In units with a grid, inferior image quality was due to the low optical density of the images, the use of inadequate image receptors and the use of an aluminium filter. Image quality results were compared with the standards of the UK Breast Screening Programme and proposed European Guidelines. It is concluded that mammography image quality might be considered as acceptable when TOR(MAX) phantom films achieved at least 10 lp mm-1 and 1.2% for high contrast resolution and threshold contrast visibility (6 mm details) respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.