Abstract

The aim of this study is to examine the corneal tangent modulus measurement repeatability and performance of the corneal indentation device (CID). Twenty enucleated porcine eyes were measured and the eyes were pressurized using saline solution-filled manometer to 15 and 30mmHg. Corneal tangent moduli measured using the CID were compared with those measured using high precision universal testing machine (UTM). The within-subject standard deviation (Sw), repeatability (2.77×Sw), coefficient of variation (CV) (Sw/overall mean), and intraclass correlation coefficient (ICC) were determined. The mean corneal tangent moduli measured using UTM and CID were 0.094±0.030 and 0.094±0.028MPa at 15mmHg, and 0.207±0.056 and 0.207±0.055MPa at 30mmHg, respectively, with a difference less than 0.13%. The 95% limit of agreement was between −0.009 and 0.009MPa. The Sw, repeatability, CV and ICC of corneal tangent moduli measured by the CID were 0.006MPa, 0.015MPa, 4.3% and 0.993, respectively. The results showed that the corneal tangent moduli measured by the CID are repeatable and are in good agreement with the results measured by the high precision UTM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.