Abstract

Two methods for assessing accommodative facility were compared: the Computer Orthoptics Diagnostic Program and the standard Lens Flipper Method. In random fashion, 40 visually normal [modified clinical technique (MCT) screening] subjects, ages 10 to 18 years, were tested monocularly and binocularly on both methods. Scatter-plot comparisons of the data showed no apparent relation between the two methods, with correlation coefficients close to zero for both monocular (r = 0.0205) and binocular (r = 0.1180) results. In addition, the subject's diagnostic classification (pass or fail) was compared between the two methods. The kappa statistic indicated a low level of agreement between the two methods both monocularly and binocularly. These results suggest that the Computer Orthoptics' diagnostic test is not a valid method for diagnosing accommodative facility deficiencies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.