Abstract
Objective To compare corneal thickness (CT) measurements obtained by specular microscopy,Slit-scanning pachymetry,anterior segment optical coherence tomography (AS-OCT) and ultrasound pachymetry,with a focus on central and midpeipheral (3.0 mm) region of cornea. Methods Central CT and CT at four midperipheral points were measured with the specular microscopy (Topcon SP-3000P),Slit-scanning pachymetry (Orbscan),AS-OCT and ultrasound pachymetry. Results In 134 eyes of 134 healthy subjects,the mean central CT was 543.10±32.69μm for ultrasound pachymetry,537.02±32.39 μm for AS-OCT,534.93±34.25μtm for Orbsan Ⅱ and 515.49±32.76μm for specular microscopy.No significant difference was found between AS-OCT measurements and Orbsan Ⅱ (P =0.07).The best agreement was observed in the ultrasound pachymetry and AS-OCT.In 125 eyes,the measurements of mean midperipheral CT obtained by Orbsan Ⅱ were much higher than AS-OCT and specular microscopy at all points. Conclusions Specular microscopy measurements of central CT and midperipheral CT are much thinner than other methods.Central CT measurements obtained by Orbscan Ⅱ and AS-OCT show good agreement,therefore can be used interchangcably; but midperipheral CT measurements cannot be used interchangeably due to the large difference. Key words: Corneal thickness; Specular microscopy; Anterior segment OCT; Orbscan Ⅱ; Ultrasound pachymetry
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.