Abstract
The study compared the diagnostic performance of two shear-wave elastography techniques, elastography point quantification (ElastPQ) and 2-D shear-wave elastography by General Electric (2-D-SWE.GE), for the non-invasive assessment of liver fibrosis in a cohort of patients with chronic hepatitis C virus (HCV) infection, using transient elastography (TE) as the reference method. There was no significant difference between the feasibility rates of TE, ElastPQ and 2-D-SWE.GE (p = 0.507). A good correlation was found between the liver stiffness (LS) values obtained using the two elastographic methods (r = 0.78). The mean LS values obtained using the ElastPQ technique were significantly higher than those obtained using 2-D-SWE.GE (12.1 ± 7.3 kPa vs. 10.4 ± 4.0 kPa, p < 0.0001). Pairwise comparisons of receiver operator characteristics curves between 2-D-SWE.GE and ElastPQ have shown that there are no significant differences in their performance for staging F ≥ 2 fibrosis (p = 0.89), F ≥ 3 fibrosis (p = 0.76) and F = 4 fibrosis (p = 0.86) in patients with chronic HCV infection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.