Abstract
The performance of two capillary electrophoresis (CE) instruments, one commercial and one homemade device, were compared for the determination of derivatised aminated compounds with fluorescein isothiocyanate (FITC). The commercial CE system first uses an argon ion laser as excitation source; the homemade CE device uses an inexpensive blue-light-emitting diode (LED) as the light source and a charge-coupled device (CCD) as the detection system. After fine optimisation of several separation parameters in both devices, a co-electroosmotic flow CE methodology was achieved in coated capillary tubing with 0.001% hexadimetrine bromide (HDB), and 50 mmol L-1 sodium borate at pH 9.3 with 20% 2-propanol for the determination of several amines and aminoacids. Analytical performances, applicability in beer samples and other aspects such as cost or potential for miniaturization have been compared for both devices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.