Abstract

This article, for comparison, the non-proportionality of light yield and energy resolution of BGO, LYSO and CsI(Tl) scintillators couple to the R1306 PMT readouts were investigated. At 662 keV from 137Cs source, the good energy resolution of 7.13% for CsI(Tl) superior than LYSO and BGO scintillators. The energy resolution on gamma-ray energy was also evaluated to expose the scintillator intrinsic resolution parameters. For non-proportionality of light yield, the study showed a light yield non-proportionality 0.35% of LYSO, the value is better than 4.82 % for CsI(Tl) and 1.53 % of BGO scintillators.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.