Abstract

A new, commercially available roller pump system able to deliver pulsatile and nonpulsatile flow has been studied in patients undergoing elective open-heart surgical procedures. The pulsatile pump (Stöckert Instrumente) may be used with standard extracorporeal circuit equipment and consistently produces a peripheral arterial pulse pressure of 25 to 30 mm. Hg at mean flow rates of 3.5 to 4.0 L. per minute. Twenty patients, arbitrarily allocated to pulsatile or nonpulsatile groups, have been studied. There were no significant differences between the groups in respect of age, weight, bypass time, pump flow, or mean arterial pressure during bypass. Comparative studies of the hematologic effects of pulsatile and nonpulsatile perfusion were carried out. There was no evidence of increased hemolysis with pulsatile flow, nor was there increased depletion of red blood cells (RBC's) or platelets in the pulsatile group. This pulsatile pump system may therefore be used to produce pulsatile perfusion during cardiopulmonary bypass without the fear of producing excessive blood cell trauma.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.