Abstract

Graphene has been employed as electrode materials in various electrochemical biosensors due to its excellent electrical, mechanical, thermal and optical properties. In the present study, Chemical Vapor Deposited (CVD) and epitaxial graphene on SiC were examined as material for electrochemical biosensing application. The surface of both types of graphene were characterized using Raman spectroscopy as well as with Scanning Electron Microscopy (SEM). As the key point for the comparison, the impedance spectroscopy measurements of different graphene films using deionized water and saline 0.9% NaCl solution were performed as well. The method of impedance measurements applied to graphene films expands the range of possibilities for using this material as sensitive biosensors. Based on the comparative tests results, it is possible to draw the first simple conclusions about the advantages of CVD or epitaxial graphene. Based on the results of impedance spectroscopy, it is possible to draw a simple conclusion – single layer graphene has the higher sensitivity.