Abstract

To compare the parameters of three different topographic devices (Pentacam HR, Oculus Optikgeräte GmbH; Sirius, Costruzione Strumenti Oftalmici; and Cassini, i-Optics) in grading the severity of keratoconus in cross-linked and non-cross-linked eyes. This was a prospective comparative interventional study done in a tertiary eye care center, wherein 114 eyes of 68 patients with keratoconus were divided into two groups: 62 eyes that were observed and 52 eyes that were cross-linked. All eyes were evaluated on all three topographers at baseline, 3-month follow-up, and 6-month follow-up. The Sirius showed significantly lower values of mean flat keratometry in comparison to the Pentacam HR (limits of agreement [LoA]: 1.75 to 3.51%) and Cassini (LoA: 1.75 to 3.51%). The mean steep keratometry values were higher for the Cassini in comparison to the Pentacam HR (LoA: 4.39 to 7.02%) and Sirius (LoA: 3.51 to 6.14%). The mean steep keratometry values of the Sirius were significantly lower than those of the Pentacam HR and in both the cross-linked and observation groups (LoA: 3.51 to 6.14%). The mean keratometry values were significantly higher in the Cassini compared to the Pentacam HR (LoA: 3.51 to 8.77%) and significantly lower in the Sirius in comparison to the Pentacam HR (LoA: 3.51 to 6.14%). The mean difference in astigmatism was also statistically significant between the three tomographers in both groups, with the Cassini showing higher values (LoA: 3.51 to 5.26%) and the Sirius lower values than the Pentacam HR (LoA: 5.26 to 6.14%). The authors conclude the three devices cannot be used interchangeably. The Cassini showed better agreement and correlation with the Pentacam HR in the staging of keratoconus, whereas the Sirius tended to underdiagnose and under-stage the disease. [J Refract Surg. 2024;40(4):e260-e269.].

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call