Abstract

Quantification of human immunodeficiency virus type 1 (HIV-1) RNA in plasma has become the standard of care in the management of HIV-infected patients. There are several commercially available assays that have been implemented for the detection of HIV-1 RNA in plasma. Here, the new Hologic Aptima® HIV-1 Quant Dx assay (Aptima HIV) was compared to the Roche COBAS® TaqMan® HIV-1 Test v2.0 for use with the High Pure System (HPS/CTM). The performance characteristics of the assays were assessed using commercially available HIV reference panels, dilution of the WHO 3rd International HIV-1 RNA International Standard (WHO-IS) and plasma from clinical specimens. Assay performance was determined by linear regression, Deming correlation analysis and Bland-Altman analysis. Testing of HIV-1 reference panels revealed excellent agreement. The 61 clinical specimens quantified in both assays were linearly associated and strongly correlated. The Aptima HIV assay offers performance comparable to that of the HPS/CTM assay and, as it is run on a fully automated platform, a significantly improved workflow.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call