Abstract

Background: Welding emits harmful rays to the eyes. We evaluate and compare the UV, blue light and IR transmittance characteristics of some currently available welding safety protectors with ANSI Z87.1 criteria. Materials and Methods: Three type of welding safety protectors (three of each type) have been evaluated for spectral transmittance. One-sample T-test was performed to establish is there a statistically significant difference between the standard criteria and UV, blue and IR regions for the welding protectors (α=0.05). Results: In ultraviolet (UV) (far and near) region, two types of the tested protectors (P1, P2) transmitted lower than the maximum allowable value specified in the standard (P<0.001). In infrared (IR) spectrum only one type (P3) transmitted lower than the maximum allowable value specified in the standard (P<0.001). In blue light region, all of the tested protectors transmitted lower than the maximum allowable value specified in the standard (P<0.001). Conclusion: Although all of the tested glasses could meet some parts of the criteria, but none of them could meet the complete spectral transmittance requirements of ANSI Z87.1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.